PN B05577AC
6-67
Sample Modules
Universal Liquid Module
6
ULM Description
The ULM system consists of:
•
The ULM module
•
Two hoses: one for fluid delivery and another for waste.
The ULM requires relatively small volumes of sample and dispersing liquid (total volume 125 mL). It
is compatible with both organic and aqueous media. The set point for the obscuration is user
selectable or may be attained automatically as the sample is added to the vessel.
This sample module has the capability of auto-filling and auto-rinsing using a self-contained
internal pump. The pump is controlled by an automated level sensor that signals when the
container is full or empty. The pump fills and rinses the unit automatically, expelling all waste into
a waste container (provided with the module).
Loading the ULM
1
Open the LS 13 320 door to load the ULM by undocking the auto-docking tray. Press the
OPEN
button on the instrument and the sliding door moves to the left. When it is open, press the
EJECT
MODULE
button to extend the tray toward the user.
2
In the auto-dock tray, you will notice stainless steel pins extending into the center tray area
both at the back of the tray and in the front of the tray. These pins fit into slots on the ULM. The
back pins prevent the module from tipping out and the forward pins secure the module in the
tray.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......