PN B05577AC
2-20
Installation
Making Measurements
To create a reference background:
1
Perform a sample analysis. If the background is within the recommended value (less than
2 x 10
6
) you may select this file as your reference background.
2
From the
Run
menu select
Run Cycle Options
(
).
3
Select
New Reference Background
.
4
From the dialog box select the file you want to use as the reference background file. Click
Open
.
The background from the file you select will then be saved in the calibration files folder using
the following filename format: ZXBYYYY.$ls, where:
X = L for the ALM
U for the ULM
M for the MLM
P for the Tornado DPS
Y = the last four digits of the optical LS 13 320 bench serial number
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......