61
Select Meter:
This parameter allows you to select the meter orientation during testing.
Step Name:
This parameter allows you to enter a text name for each step.
Leakage:
This parameter allows you to measure leakage using either Peak or RMS
measurements.
Neutral:
This parameter controls the Neutral relay during Leakage Current testing.
Reverse:
This parameter controls the Reverse relay during Leakage Current testing.
Ground:
This parameter controls the Ground relay during Leakage Current testing.
Measure Device:
This parameter allows you to select the MD used during Leakage
Current testing.
Probe:
This parameter allows you to set the probe configuration during Leakage
Current testing.
AC/DC:
The metered value of leakage current during a Leakage Current test. Can be
set to AC+DC, AC or DC. When set to AC+DC, the unit measures the composite
leakage current through the MD. When set to AC, the unit measures only the AC
component of the leakage current through the MD. When set to DC, the unit measures
only the DC component of the leakage current through the MD*.
NOTE:
If the AC/DC parameter is set to DC or AC, the minimum delay time is
automatically set to a minimum of 1.3sec. If the AC/DC parameter is set to
AC+DC, the minimum delay time is automatically set to a minimum of 1.8sec.
Ranging Mode:
Unit ranging setting used during the Leakage Current test. Can be
set to Auto or Manual.
Defaults
: Pressing this soft key will default the test parameters to their predetermined
default values (refer to section
4.4.2. Default Parameters
). Once the default soft key
is pressed it is necessary to complete the edit either by pressing the ENTER key to
accept the parameter overwrite or the EXIT key to escape from the edit and return to
the original values.
DUT Output (82x6 and 82x7):
This function enables or disables high voltage at the
DUT outputs during an AC Withstand, DC Withstand, or IR test. The DUT Output
parameter will only be present if the DUT-HV Setup is selected ON at the Hardware
screen. The DUT Output parameter will appear on the AC Withstand, DC Withstand,
or IR test parameter screens as follows:
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...