151
After the barcodes are scanned, pressing TEST will initiate the test sequence.
Pressing RESET will abort the TEST sequence.
The Data Storage Card allows for the re-scanning of barcodes if the previously
scanned barcode was incorrect. Re-scanning is available in the SERIAL#,
PRODUCT# and SER/PROD modes. Anytime before a test is initiated, the user
can re-scan a barcode. If the user decides to re-scan barcodes when the
Barcode I/P setting is set to SER/PROD, the Data Storage Card will first replace
the data in the Serial Number field, and if the user re-scans another barcode, the
Data Storage Card will replace the data in the Product Number field.
The RUN FILE selection gives the user the ability to automatically load and
execute a test file based on what barcode is scanned from the Perform Tests
screen.
Using the RUN FILE feature will enable the instrument’s
output once the barcode is scanned. Do not touch the DUT
at any time when using this feature in order to avoid
potential shock or serious injury.
To completely enable this feature, the user must name the desired test file for a
particular product the exact alpha-
numeric code that is on the product’s barcode
label. For example, if Product A has barcode “123456789”, then the test file that
the user would like to run when testing Product A should be n
amed “123456789”.
Upon scanning the barcode, the OMNIA II will immediately execute the test
associated with that barcode. The test file name is limited to 10 characters.
However, if the user names a test file with the maximum 10 characters, this
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...