152
functio
n will still initiate a test when a product’s barcode begins with those first 10
characters even if the barcode has more than 10 characters.
When using certain features of barcoding, the OMNIA II
’s response to the TD?
and RD x? commands will differ slightly to an OMNIA II with the standard RS232
interface. For all types of tests (ACW, DCW, IR, GND, CONT, LCT, RUN) two
fields are added to the end of the standard response when the Barcode I/P
setting is set to SERIAL#, PRODUCT# or SER/PROD. The first field contains
the Serial Number information and the second field includes the Product Number
information. Both fields are included regardless of which of these three modes
are selected. The Data Storage Card
will simply substitute a “0” for the field if it
is not applicable to the setting. For example, if a user had their Barcode I/P
setting set to SERIAL#, and scanned a Serial Number with the value
“123456789”, the TD? response for an ACW test could be:
01,ACW,Pass,1.24,1.000,0.900,1.0,123456789,0
Note that there is a “0” in the Product Number field because the Barcode I/P
setting is SERIAL#.
When the Barcode I/P setting is RUN FILE or OFF, these fields are not included
in the TD? and RD x? responses.
Operation
The Data Storage Card Results menu can be viewed by pressing the Results soft
key while in the Perform Tests Screen. After doing so the following menu will be
shown:
Pressing the soft key labeled Instrument will display the detailed results from the
previously executed test file. Pressing the soft key labeled Data Card will display
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...