65
4.5. Setting Up a Test
Upon power up, OMNIA will initialize with the test file that was last used before power
down. If the instrument is new it will come loaded with one default file that the
instrument will boot up with until it is changed.
1. From the Setup Test
s screen, press the “File” soft key. The File Setup screen
will now be displayed.
2. From the File Setup
screen, press the soft key “New File”. The Create File
screen will now be displayed.
3. The Create File screen is separated into two main sections: File Name Edit and
Soft Key Menu.
4. From the Create File screen you must give your new test file a name. The file
name can be a maximum of fifteen characters in length and may use any
combination of the character set plus numbers and symbols. To name your
file, use the alphanumeric keypad. If you make a mistake or want to change
the character, press the backspace key in the alphanumeric keypad. If you wish
to use capital letters in
the prompt, press the “Caps Lock” soft key. While
pressing the Caps Lock soft key, the Caps Lock parameter will toggle between
ON and OFF. If you wish to use symbols in the prompt, press the “Symbol Key”
soft key. While pressing the Symbol Key soft key, the symbol values will toggle
between !, *, _, - and ~. When you have finished editing the file name press the
ENTER key.
5. From the Setup Test
s screen, press the “Add” soft key. The Test Selection
screen will now be displayed. From this screen you may choose what type of
test to perform: Dielectric Withstand, Insulation Resistance, AC Ground Bond
or DC Continuity.
6. From the Test Selection screen, press the soft key next to the type of test you
wish to add to the test sequence.
If you press the “Dielectric Withstand” soft
key, the AC-Withstand and DC-Withstand test soft keys will be displayed. Press
the AC-Withstand or DC-Withstand soft key to choose a dielectric withstand
test type. The Parameter Setting screen for the type of test you have chosen
will now be displayed.
7. From Parameter Setting screen you may set your own custom parameters for
the test or choose the defaults from the soft key menu. All of the individual
parameters for the test may be accessed using the up and down arrow keys,
and the ENTER key. As the arrow keys are pressed, the highlighted area will
scroll to the different parameters. The ENTER key may also be used to scroll
to the different parameters. At the bottom of the screen is the Scanner Setup
parameter (for a detailed explanation of how this feature works, refer to the
section
5.1.5. Scanner Connections
).
8. Scroll to a parameter you wish to edit. The parameter that is highlighted may
now be edited (refer to section
4.4. Test Parameters
). Enter a new number via
the alphanumeric keypad. Once you begin typing a new number the parameter
will blank and the cursor will begin blinking. This indicates that the parameter is
being edited. Once a parameter is edited it is necessary to complete the edit
either by pressing the ENTER key to accept the new number or the EXIT key to
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...