GROUND BOND TEST MODE
Output Voltage
(Open Circuit Limit)
Range:
3.00
– 8.00 VAC
Resolution: 0.01 V
Accuracy:
(2 % of s 0.03V) O.C. Condition
Output Frequency
Range:
60 or 50 Hz, user selectable
Accuracy:
0.1%
Output Current
Range:
1.00
– 40.00 A
Resolution: 0.01 A
Accuracy:
(2 % of s 0.02 A)
Output Regulation
Accuracy:
(1% of 0.02A)
Within maximum load limits, and over input voltage range.
Maximum Loading
1.00
– 10.00A, 0 – 600m
10.01
– 30.00A, 0 – 200m
30.01
– 40.00A, 0 – 150m
Current Display
Range:
0.00
– 40.00 A
Resolution: 0.01 A
Accuracy:
(3 % of s 0.03 A)
HI and LO-Limits
Range:
0
– 150 m
for 30.01
– 40.00 Amps
0
– 200 m
for 10.01
– 30.00 Amps
0
– 600 m
for 1.00
– 10.00 Amps
Resolution: 1 m
Accuracy:
Same as Ohmmeter Display
Ohmmeter Display
Range:
0
– 150 m
for 30.01
– 40.00 Amps
0
– 200 m
for 10.01
– 30.00 Amps
0
– 600 m
for 6.00
– 10.00 Amps
Resolution: 1 m
Accuracy:
(2
of reading
+
2 m
)
Range:
0
– 600 m
for 1.00
– 5.99 Amps
Resolution: 1 m
Accuracy:
(3
of reading
+
3 m
)
Dwell Timer
Range:
0.5
– 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy:
(0.1% + 0.05 sec)
Milliohm Offset
Range:
0
– 200m
Resolution: 1 m
Accuracy:
(2 % of s 2 m
)
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...