![Associated Research OMNIA II 8204 Operation And Service Manual Download Page 245](http://html1.mh-extra.com/html/associated-research/omnia-ii-8204/omnia-ii-8204_operation-and-service-manual_2989483245.webp)
239
Range 4
400uA ~ 999.9uA, 1000uA-2100uA,
frequency DC, 15Hz - 1MHz
Resolution
0.1uA, 1uA
Accuracy
15Hz < f <1MHz: ±(10% of r
2uA)
Range 5
1800uA ~ 8400uA, frequency DC,
15Hz - 1MHz
Resolution
1uA
Accuracy
15Hz < f <1MHz: ±(10% of r
2uA)
Range 6
8000uA ~ 8399uA, 8.40mA ~
10.00mA, frequency DC, 15Hz
–
100KHz
Resolution
1uA, 0.01mA
Accuracy
15Hz < f <100KHz: ±(10% of r
2counts)
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display (Peak) (AC
Only)
Range 1
0.0uA ~ 32.0uA, frequency 15Hz
- 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Range 2
7.8uA ~ 130.0uA, frequency 15Hz
- 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Range 3
31.5uA ~ 525.0uA, frequency 15Hz
- 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA, frequency 15Hz
- 1MHz
Resolution
1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...