247
Additional Error
15Hz < f < 30Hz : add ±(2% of reading
+ 0.2% of range) to base accuracy
30Hz < f < 100KHz: add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 6
0.0mV ~ 600mV, frequency DC
–
100KHz
Resolution
1mV
Additional Error
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
DISPLAY AUTO RANGE
Touch Voltage Display (DC Only)
Range 1
0.0mV ~ 32.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 2
28.0mV ~ 130.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 3
120.0mV ~ 525.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 4
400mV ~ 2100mV, DC
Resolution
0.1mV, 1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 5
1800mV ~ 8400mV, DC
Resolution
1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 6
8.00V ~ 10.00V, DC
Resolution
1mV, 0.01V
Accuracy
DC : ±5% of reading (0.01V -10.00V )
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Voltage Display (DC Only)
Range 1
0.0mV ~ 32.0mV, DC
Summary of Contents for OMNIA II 8204