126
COMMAND
NAME
VALUE
SCA {1|0}
SCA?
Cal Alert
1= On, 0=Off
COMMAND
NAME
VALUE
SCDA <
value
>
SCDA?
Cal Date
mm,dd,yy or yy,mm,dd or
dd,mm,yy ; same as SDF
setting
SCDU <
value
>
SCDU?
Cal Due
mm,dd,yy or yy,mm,dd or
dd,mm,yy ; same as SDF
setting
SDAY {6|5|4|3|2|1|0}
SDAY?
Day of the week
0,1,2,3,4,5,6=
Sun,M,T,W,T,F,Sat
SDF <
value
>
SDF?
Date Format
0=yy,mm,dd, 1=mm,dd,yy
2=dd,mm,yy
SDH{1|0}
SDH?
DUT-HV
1= On, 0=Off
SDS
SDS?
System Continuity
Scanner
0=HV Channel
1=GND Channel
SDT <
value
>
SDT?
Date
mm,dd,yy or yy,mm,dd or
dd,mm,yy ; same as SDF
setting
SF {1|0}
SF?
Fail Stop
1= On, 0=Off
SL {1|0}
SL?
Lock
1= On, 0=Off
SM {1|0}
SM?
System Measurement
0 = True RMS, 1 = Average
SMM {2|1|0}
SMM?
Main Menu page
2 = My Menu, 1= Perform
Tests, 0=Menu
SPR {1|0}
SPR?
PLC Remote
1= On, 0=Off
SSG {1|0}
SSG?
Smart GFI
1= On, 0=Off
SSG2 < value >
SSG2?
Smart GFI value
0.4
– 5.0, 0
SSI {1|0}
SSI?
Single Step
1= On, 0=Off
STM <
value
>
STM?
Time
hh,mm (24hr)
hh,mm,AM or hh,mm,PM (12hr)
according to STF setting
STF {1|0}
STF?
Time Format
0=12hr, 1=24hr
CAD {1|0}
CAD?
System Alert Date
1 = On, 0 = Off
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...