118
ADD IR,1000,0.00,0.05,0.1,1.0,0.0,0.000,OFF,OOOOOOOO
Adding a Ground Bond Test
ADD GND,Current,Voltage,HI-Limit,LO-Limit,Dwell Time,Offset,Scanner
Channel,Frequency
ADD GND,25.00,8.00,100,0,1.0,0,0,60
Adding a Continuity Test
ADD CONT.,HI-Limit,LO-Limit,Dwell,Offset,Scanner Channel,Continuity
ADD CONT.,1000,0.00,1.0,0.00,OOOOOOOO
Adding a Functional Run Test
ADD RUN,Voltage-HI,Voltage-LO,Amp-HI,Amp-LO,Dwell,Delay,Leakage-HI,Leakage-
LO,Power-HI,Power-LO,PF-HI,PF-LO,Active Link,PLC Control
ADD RUN,125.0,0.0,10.00,0.00,1.0,0.2,10.00,0.00,1000,0,1.000,0.000,M1
Adding a Leakage Current test
ADD LCT,Leakage-HI,Leakage-LO,Voltage-HI,Voltage-
LO,Delay,Neutral,Reverse,Ground,Meas. Device,Probe,Peak-RMS,PLC Control
ADD LCT,6000,0.0,125.0,0.0,1.0,CLOSED,OFF,CLOSED,UL544NP,Ground To
Line,RMS,M1
ADD2
This command inserts or adds a new step to the test sequence. This command will
add or insert the test at the step location that has been selected. When a test has
already been created at a selected step then the new step will be inserted and the
previous step will move down to the next step and all other step will be renumbered
accordingly.
The parameter <
test>
indicated the test type. The values ACW, DCW, IR, GND,
CONT, RUN or LCT must be used. The parameters <
p1,p2>
etc. indicate the
individual settings for each test. All parameters must be included with the command
and should appear in the same order that is shown on the actual setting screens. The
correct order is shown in the tables below. When the scanner is installed it should be
the last parameter for all test types except GND. For the GND test type, the scanner
settings should be inserted between offset and frequency. The list of parameters can
also be found in the default parameters section of the manual.
The parameter values should use complete text and not use the coded values that are
associated with the individual parameter setting commands. Such as "ON" and "OFF"
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...