17 GB Voltage (Discontinued)
The GB Voltage Option allows user to set up Voltage Hi-Limit and Lo-Limit during AC
Ground Bond test. During the test the Omnia II’s display shows a Voltage meter that
reflects the amount of Voltage dropped across the device under test.
From the Setup Tests Selection screen, press the
“Add” soft key. The test types soft
keys will now be displayed. From the Setup Tests Selection screen,
press the “AC
Ground Bond
” soft key. The Ground Bond Parameter Setting screen will now be
displayed. The Ground Bond Parameter Setting screen will appear as follows:
From the Ground Bond Parameter Setting screen, the following parameters may be
controlled: Current, Voltage, HI-Limit, LO- Limit, Dwell Time, Offset, Scanner Channel,
Frequency, DualCHEK, Step Name and Defaults.
Offset
This function allows the instrument to compensate for lead and test fixture voltage
drop during a Ground Bond. Using the up and down arrow keys or the ENTER key,
scroll the highlighted area to the Offset parameter. You may now manually or
automatically set an Offset value.
To manually set an Offset value enter a Volt value via the numeric keypad and then
press the ENTER key to accept the new value or press the EXIT key to escape from
the edit.
To automatically set an Offset value set the output voltage, current, and frequency to
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...