109
If the Remote Interlock contacts are opened during a test, the pop-up message will be
displayed and the test will abort. The hardware and has been configured to provide
the interlock connections on pins 4 and 5 of the Remote Interface, Signal Input port.
The instrument can still be used without the external interlock device as long as the
Interlock Disable Key (38075 provided with unit) is plugged into the Remote Interface,
Signal Input port. If there is nothing connected to the Remote Interface, Signal Input
port to provide a connection to the Remote Interlock, the instrument will not perform
tests.