This message appears on the display if the DUT input voltage drops below the Volt-
LO Trip setting.
Amp-HI
This message appears on the display if the current exceeds the Amp-HI Trip setting.
Amp-LO
This message appears on the display if the current drops below the Amp-LO Trip
setting.
Line-OC
This message appears on the display if the 20 Amp < 1 s, short circuit protection is
tripped.
Watt-HI
This message appears on the display if the DUT input wattage exceeds the Watt-HI
Trip setting.
Watt-LO
This message appears on the display if the DUT input wattage drops below the Watt-
LO Trip setting.
PF-HI
This message appears on the display if the Power Factor exceeds the PF-HI Trip
setting.
PF-LO
This message appears on the display if the Power Factor drops below the PF-LO Trip
setting.
Leak-HI
This message appears on the display if the enclosure to neutral leakage current
exceeds the Leak-HI Trip setting.
Leak-LO
This message appears on the display if the enclosure to neutral leakage current drops
below the Leak-LO Trip setting.
Neutral-V
This message appears on the display if there is voltage present on the Neutral of the
DUT input receptacle.
OCP
This message appears on the display if the maximum current draw exceeds the
capability of the internal power source (>4.2A).
A-SHDN
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...