219
and enter it using the numeric keypad. You may now store the new calibration
number by pressing the ENTER key or escape by pressing the EXIT key or the
RESET button.
Calibration of LCT MD: UL 544NP Measuring Device
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect 2.0VDC in series
with a current meter to
Probe-HI + and Probe-LO
-.
When the power supply
and meter are connected,
press TEST to start the
calibration.
Enter Standard Current
Reading.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (
Cal. Active
).
You may now store the new calibration number by pressing the ENTER key or escape
by pressing the EXIT key or the RESET button.
Calibration of LCT MD: UL 544NP Measuring Device
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect 2.0VDC in series
with a current meter to
Probe-HI + and Probe-LO
-.
When the power supply
and meter are connected,
press TEST to start the
calibration.
Enter Standard Current
Reading.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (
Cal. Active
).
Read the measurement from your standard and enter it using the numeric keypad.
You may now store the new calibration number by pressing the ENTER key or escape
by pressing the EXIT key or the RESET button.
Calibration of LCT MD: IEC60601-1 / UL2601 Measuring Device
The load/meter screen and prompt screens will contain the following text for this
calibration point:
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Connect 2.0VDC in series
with a current meter to
Probe-HI + and Probe-LO
-.
When the power supply
and meter are connected,
press TEST to start the
calibration.
Enter Standard Current
Reading.
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...