DIELECTRIC WITHSTAND TEST MODE (CONT.)
Current Display
AC Total
AC Real
DC
Auto Range*
Range 1: 0.000 mA
– 3.500 mA RMS
(0.000 mA
– 5.500 mA Peak)
Resolution: 0.001 mA
Range 2
3.00 mA
– 50.00 mA RMS
(5.00 mA
– 70.07 mA Peak)
Resolution: 0.01 mA
Accuracy:
(2% of r 2 counts) All Ranges
Range 1:
0.000 mA
– 9.999 mA
Resolution: 0.001 mA
Range 2:
10.00 mA
– 50.00 mA
Resolution: 0.01 mA
Accuracy:
(3% of r 50
A) All Ranges
PF > 0.1
V > 250VAC
Range 1:
0.0
A
– 350.0
A DC
(0.0
A
– 550.0
A Peak)
Resolution: 0.1
A
Range 2 :
0.300 mA
–3.500 mA DC
(0.400 mA
–5.500 mA Peak)
Resolution: 0.001 mA
Range 3
3.00 mA
– 20.00 mA DC
(4.00 mA
– 25.00 mA Peak)
Resolution: 0.01 mA
Accuracy:
(2% of r 2 counts) All Ranges
DC Output Ripple
4% Ripple RMS at 5 KV DC @ 20mA, Resistive Load
Discharge Time
50 ms no load, < 100 ms for capacitive load
Maximum Capacitive
Load
DC Mode
1uF
< 1KV
0.75uF < 2KV
0.5uF < 3KV
0.08uF < 4KV
0.04uF < 6KV
AC Output
Waveform
Sine Wave, Crest Factor = 1.3
– 1.5
Output Frequency
Range:
60 or 50 Hz, User Selection
Accuracy:
0.1 %
Output Regulation
(1 % of 5 V)
from no load to full load and over input voltage range.
Dwell Timer
Range:
AC 0.4
–999.9 sec (0 = Continuous)
Range:
DC 0.3
–999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy:
(0.1% + 0.05 sec)
Ramp Timer
Range:
Ramp-Up:
AC 0.1
– 999.9 sec
DC 0.4
– 999.9 sec
Ramp-Down:
AC 0.0
– 999.9 sec
DC 0.0 , 1.0
– 999.9 sec
Resolution: 0.1 sec
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...