243
Basic Accuracy
DC, 15Hz < f < 100KHz : ±5% of
reading (0.01V -20.00V )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Voltage Display (RMS)
(AC+DC)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0mV ~ 7.8mV, frequency
15Hz - 1MHz
Resolution
0.1mV
Additional Error
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 3
0.0mV ~ 31.5mV, frequency
15Hz - 1MHz
Resolution
0.1mV
Additional Error
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 4
0.0mV ~ 126.0mV, frequency
15Hz - 1MHz
Resolution
1mV
Additional Error
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 5
0.0mV ~ 504mV, frequency
15Hz - 1MHz
Resolution
1mV
Additional Error
15Hz < f < 100KHz : add ±(2% of
r 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
r 0.5% of range) to base
accuracy
Range 6
0.0mV ~ 600mV, frequency DC
–
100KHz
Summary of Contents for OMNIA II 8204