245
Resolution
1mV, 0.01V
Accuracy
15Hz < f < 100KHz : ±5% of reading
(0.01V -10.00V )
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Voltage Display (RMS) (AC
Only)
Range 1
0.0mV ~ 32.0mV, frequency DC,
15Hz
– 1MHz
Resolution
0.1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of r
5counts)
30Hz < f <100KHz: ±(2% of r
3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 2
7.8mV ~ 130.0mV, frequency DC,
15Hz - 1MHz
Resolution
0.1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of r
5counts)
30Hz < f <100KHz: ±(2% of r
3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 3
31.5mV ~ 525.0mV, frequency DC,
15Hz - 1MHz
Resolution
0.1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of r
5counts)
30Hz < f <100KHz: ±(2% of r
3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 4
126.0mV ~ 999.9mV,1000mV ~
2100mV, frequency DC, 15Hz - 1MHz
Resolution
1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of r
5counts)
30Hz < f <100KHz: ±(2% of r
3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 5
504.0mV ~ 999.9mV, 1000mV ~
8400mV, frequency DC, 15Hz - 1MHz
Resolution
1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of r
5counts)
30Hz < f <100KHz: ±(2% of r
3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 6
600mV ~ 8399mV, 8.40V ~ 10.00V,
frequency DC, 15Hz
– 100kHz
Resolution
1mV, 0.01V
Summary of Contents for OMNIA II 8204
Page 8: ...2 This Page Intentionally Left Blank ...
Page 87: ...81 conditions The different combinations are represented in the table below ...
Page 91: ...85 ...
Page 92: ...86 ...
Page 139: ...133 7 128 Power On PROMPT ESR ESE STB SPOLL SRE ESE SRE ...
Page 142: ...136 ...
Page 156: ...150 appear on the display ...
Page 169: ...163 Print Report Example ...
Page 181: ...175 ...