Check for bubbles in the pump system
1.
Check for bubbles at any point where the pump channels join.
1
2
3
1
Array ferrule tip
2
All tubing
3
Elbow (internal channel) of the lower pump block
2.
If bubbles are present, run the Bubble Remove Wizard. For more information,
see “Typical conditions for using maintenance wizards“ on page 117.
IMPORTANT!
When this wizard runs, it introduces polymer into the anode buffer jar.
Replace the anode jar buffer after you run this wizard. See “Fill the anode buffer
Chapter 2
Prepare the instrument
Check for bubbles in the pump system
2
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
35