5.
Select the plate records to add, then click Add.
6.
When the needed plate records are listed, click Done.
7.
As needed, click Up, Down, or Remove to organize the list of plate records in the
Input Stack list.
8.
Load the plates in the In Stack in the order to match the corresponding plate
record. The bottom plate runs first and uses the first plate record in the list.
IMPORTANT!
The order of the plate records must match the stack order of the
plates in the In Stack. If the order does not match, data files will contain the
wrong plate record information.
Note:
You can assign more plates in the Run Scheduler than are actually
available in the stacker.
9.
Click (Run).
During a manual run, the following steps occur.
• Plates are listed in the Input Stack
• The currently running plate is listed in the Auto Sampler field
• Completed plates are listed in the Output Stack
For more information, see “Monitor the status of the run“ on page 65.
Change injection order of plate quadrants
You can change the run order of quadrants. Changing the order of the quadrants
overrides the setting in the plate record and is applied to all plates.
1.
Select GA Instruments
4
ga3730
4
Instrument name
4
Run Scheduler.
2.
At the top of the screen, select Instrument
4
Scheduling Preference.
Note:
The Instrument menu is available only when the Run Scheduler is
displayed.
3.
Select the quadrant priority (run order)
from the Quadrant list.
You can select any run order. For
information on the order of well injection,
see the following sections.
• “Default injection positions—48-
• “Default injection positions—96-
• “Default injection order on plates“ on page 59
Chapter 5
Run the instrument
Change injection order of plate quadrants
5
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide