Troubleshooting
■
Restart the instrument and the computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
■
Display the Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
■
Review a run log file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
■
Formamide preparation and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
■
Front panel status light troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
■
Instrument run troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
■
Re-extraction troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
■
Service console troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
■
Spatial calibration troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
■
Spectral calibration troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
■
General troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Restart the instrument and the computer
Use this procedure if any of the following conditions occur.
• If communication errors are displayed
• If the front panel indicator is blinking red
• At the end of spatial calibration, if Accept/Reject buttons are dimmed
• If maintenance wizards are taking longer than expected
• If software operations are taking longer than expected
When you are instructed to restart the instrument and the computer:
1.
Exit the Data Collection Software.
2.
Power off the computer.
3.
Make sure the instrument door is closed, then power off the instrument.
4.
When the computer is completely powered off, wait 60 seconds, then power on
the computer. Wait until the Windows
™
login screen is displayed. Do not log in.
5.
Power on the instrument and wait until the green status light on the front panel
is on and not flashing before proceeding.
6.
Log in to Windows
™
operating system.
7.
Start the Data Collection Software.
A
134