Store a capillary array
CAUTION!
During storage, keep both ends of the capillary array immersed in
1X run buffer. If the capillary array dries out, it may be damaged.
1.
Remove the capillary array from the instrument using the Install Array wizard.
IMPORTANT!
Do not select Discard Array if you want to install the capillary
array on the instrument again. If you click Discard, the software flags the array
information as "discarded" will not allow you to enter the array information
again.
2.
Place 80 mL of 1X run buffer in the capillary array header shipping cover.
3.
Lower the capillary tips of the array header
into the shipping cover and lock the header
onto the cover. The tips of the capillaries
should be immersed in buffer.
4.
Clip the detection cell window cover onto
the detection cell.
5.
Attach the detection cell with cover to the
storage post on the capillary array frame.
1
2
3
1
Detection cell with cover
2
Storage post on the capillary array frame
3
Vial with the array tip
1
2
1
Header shipping cover
2
Buffer reservoir
Chapter 9
Maintenance
Store a capillary array
9
124
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide