Condition
Requirement
Transport and storage
conditions
–30 to +60°C ( –22 to +140°F)
Humidity: minimum 5%, maximum 95% (average in a year
<80%)
Liquid waste collection
Dispose of the polymer, buffer, reagents and any liquid
waste as hazardous waste in compliance with local and
national regulations.
Other conditions
Ensure the room is away from any vents that could expel
particulate material on the components.
Avoid placing the instrument and computer adjacent to
heaters, cooling ducts, or in direct sunlight.
Electrical requirements
CAUTION!
Do not unpack or plug in any components until they are configured
for the proper operating voltage by the service representative.
WARNING!
For safety, the power outlet for the instrument must be accessible
at all times. In case of emergency, you must be able to immediately disconnect
the main power supply to all the equipment. Allow adequate space between the
wall and the equipment so that the power cords can be disconnected in case of
emergency.
• Dedicated line and ground between the instrument and the main electrical
service.
• Maximum power dissipation: 600 VA (not including computer and monitor)
• Mains AC line voltage tolerances must be up to ±10 percent of nominal voltage
• Instrument—AC 100–240 V ±10%, 50/60 Hz, 6.7 A, power rated 600 VA
• Maximum current—6.7 A
Device
Rated voltage
Circuit
required
Rated
frequency
Rated
power
Instrument
100–240 ±10% VAC
[1]
10 A
50/60 Hz
600 VA
Computer (desktop)
100–240 ±10% VAC
10 A
50/60 Hz
125 VA
Monitor
65 VA
[1]
If the supplied power fluctuates beyond the rated voltage, a power line regulator may be required. High or low
voltages can adversely affect the electronic components of the instrument.