DANGER!
LASER HAZARD. Under normal operating conditions, the 3730xl
DNA Analyzer is categorized as a Class 1 laser product. However, removing
the protective covers and (when applicable) defeating the interlock(s) may
result in exposure to the internal Class 3B laser. Lasers can burn the retina,
causing permanent blind spots. To ensure safe laser operation:
·
Never look directly into the laser beam.
·
Do not remove safety labels, instrument protective panels, or defeat safety
interlocks.
·
The system must be installed and maintained by a Thermo Fisher Scientific
Technical Representative.
·
Remove jewelry and other items that can reflect a laser beam into your eyes
or those of others
·
Wear proper eye protection and post a laser warning sign at the entrance to
the laboratory if the laser protection is defeated for servicing
·
DO NOT operate the laser when it cannot be cooled by its cooling fan; an
overheated laser can cause severe burns on contact.
The following table lists laser safety symbols and alerts that may be present on the
instrument.
Alert
DANGER!
Class 3B (III) visible and/or invisible laser radiation present when open
and interlocks defeated. Avoid exposure to beam.
CAUTION!
LASER HAZARD, Bar Code Scanner. The bar code scanner included
with the instrument system is a Class 2 laser. To avoid damage to eyes, do not
stare directly into the beam or point into another person's eyes.
Alerte (français)
DANGER !
Rayonnement laser visible ou invisible de classe 3B (III) présent en
position ouverte et avec les dispositifs de sécurité non enclenchés. Éviter toute
exposition au faisceau.
MISE EN GARDE !
RISQUE LIÉ AU RAYONNEMENT LASER, Lecteur de code-
barres. Le lecteur de code-barres inclut dans l'instrument est un appareil laser
de classe 2. Pour éviter toute lésion oculaire, ne regardez pas directement le
faisceau et ne le dirigez pas vers les yeux d'une autre personne.
Safety and electromagnetic compatibility (EMC) standards
The instrument design and manufacture complies with the standards and
requirements for safety and electromagnetic compatibility as noted in the following
table:
Laser safety
Appendix F
Safety
Safety and electromagnetic compatibility (EMC) standards
F
160
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide