Perform spectral calibration for sequencing and fragment analysis
A spectral calibration creates a matrix that is used during a run to reduce raw data
from the instrument to the 4- or 5-dye data stored in the sample files.
Note:
The run times listed below do not include the time for the oven to preheat.
Application
Standard
Polymer
Capillary array
length
Run module
Run Time (min)
[1]
Sequencing
Sequencing
standard
POP-7
™
36 cm
Spect36_SeqStd_POP7
60
50 cm
Spect50_SeqStd_POP7
115
POP-6
™
36 cm
Spect36_SeqStd_POP6
65
50 cm
Spect50_SeqStd_POP6
145
Fragment
analysis
Matrix
standard
POP-7
™
36 cm
Spect36_MtxStd_POP7
50
50 cm
Spect50_MtxStd_POP7
60
[1]
The software may take up to 30 minutes to calculate the matrices after the run is complete.
• For each dye set–capillary array type–capillary array length combination
• After the laser or CCD camera has been realigned/replaced by a service engineer
• After the capillary array is changed or replaced
• If you see a decrease in spectral separation (pull-up and/or pull-down peaks)
• If you alter any condition (dye set, array type, array length, or polymer type)
Spectral
calibration
definition and run
time
When to perform
the calibration