Dye Set
Quality Value minimum
Default Condition Number
Range
Any4Dye-HDR
>0.8
1.0–20.0
Any5Dye
IMPORTANT!
Review and evaluate the spectral calibration profile for each capillary,
even if the Q value and Condition numbers are within range.
1.
In the navigation pane of the Data Collection Software, select GA
Instruments
4
ga3730
4
Instrument name
4
Spectral Viewer.
1
2
3
1
Spectral profile
2
Raw data (matrix standards)
3
Rename or set the active calibration
2.
In the Dye Set drop-down list, select the dye set.
Evaluate the
spectral
calibration data
Chapter 8
Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8