Monitor the status of the run
1.
Select GA Instruments
4
ga3730
4
Instrument name
4
Run Scheduler.
3
5
1
2
4
1
Input stack—In manual run mode, lists the plate records for the plates to be run. In auto run mode, lists "Unknown".
2
Output stack—Validated plates. Plates that pass validation are listed as OK and will be run. Plates that fail validation are
listed as Validation failed and will not be run.
3
Autosampler—The plate currently in the autosampler. The runs for the current plate are listed under Current Runs.
4
Current runs—All injections set up for the plate in the plate record
5
Plate—The wells to be sampled when the selected injection is run are
.
For more information, see “Default injection positions—48-capillary arrays“ on page 37 and “Default injection positions—
96-capillary arrays“ on page 39.
Status
Description
Input stack
Completed
Run data for all injections in the plate record is added to the database and
autoextraction of data files is starting.
Processed
Data files for all injections in the plate record are autoextracted from the
database to the location specified in the results group.
A processed plate cannot be run again. To run the plate again, add samples
to the plate record. See “Add reinjections to a plate record“ on page 66.
Pending
Plate is waiting to be validated. The software checks that the settings in the
plate record match the settings on the instrument.
Output stack
OK
Plate has passed validation and will be run.
Validation failed
Plate has been validated and will not be run.
For more information, see the following sections.
• “Display the Event Log“ on page 135
• “Plate validation fails when you start a run“ on page 137
Chapter 5
Run the instrument
Monitor the status of the run
5
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide