Component
Specification
Internal barcode reader
Supports the following formats:
• Code 128
• Code 39
• Code 93
• LOGMARS
• EAN-8
Note:
All Applied Biosystems
™
barcoded plates for
the instrument use code 128 format.
Minimum computer requirements
• Hardware: OptiPlex
™
XE2, with Intel
™
Core
I7-47705, 3.1 GHz Processor
• Operating system: Windows
™
10 Enterprise
2016 LTSB, 64-bit (requires 5 partitions)
• Installed RAM: 16 GB
• Hard drive: 2 x 500GB SATA 3.0Gb/s and 8MB
Data Burst Cache
Environmental requirements
Condition
Requirement
Installation site
Indoor use only
Altitude
Safety tested up to 2,000 m (6,562 ft)
Electromagnetic
interference
Do not use this device in close proximity to sources of
strong electromagnetic radiation (for example, unshielded
intentional RF sources). Strong electromagnetic radiation
may interfere with the proper operation of the device.
Transient/overvoltage
category
Installation categories II
Vibration
The instrument is not adjacent to strong vibration sources,
such as a centrifuge, pump, or compressor. Excessive
vibration will affect instrument performance.
Pollution degree
II
Install the instrument in an environment that has
nonconductive pollutants such as dust particles or wood
chips. Typical environments with a Pollution Degree II
rating are laboratories and sales and commercial areas.
Operating conditions
• 15–30°C (59–86°F) (Room temperature should not
fluctuate ±2°C during an instrument run)
• 20–80% relative humidity, noncondensing
Appendix B
Instrument specifications
Environmental requirements
B
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide