82
017-882-B0-001 Rev. E (06/2018)
3.0 Operation, Continued
Entering Battery Mhos:
To enter the battery MHOS reading, press the BATT
softkey on the XM3 main menu. Next select the BATT
CONFIG sub-menu. Scroll the screen using the arrow
keys until BATT MHOS appears at the top of the screen.
Press the ENTR softkey. The APP card will now take
over control of the display and show a list of installed
batteries.
Use the softkeys to select the appropriate battery.
To enter the MHOS value, press the ENTR softkey. The
following screen will appear.
Use the arrow keys to select the flashing digit value.
Press the ENTR softkey to move to the next digit, and
so on. When all digits have been entered, the units will
flash. The display will show the entered mhos value and
the temperature compensated value. The PTS probe
temperature is used for temperature compensation.
Press the ENTR softkey one more time to save the
value.
Only the temperature compensated value will be logged
or displayed from this point forward.
B
ATT A1 MHOS
0M
BATT A2 MHOS
0M
BATT A3 MHOS
0M
ENTR
ESC
BATT A1 MHOS
0000M
ADJUST VALUE
ENTR TO SHIFT FIELD
ENTR
ESC
BATT A1 MHOS
1200
M
@ 75F / 24C = 1209
ENTR
ESC