N9360A Multi UE Tester GSM Programming Manual
5
Programming Command Reference
Power Ramp Trace Number
TXANalyzer:RESults:RAMP:TRACe:NUMber <int>
TXANalyzer:RESults:RAMP:TRACe:NUMber?
This command sets the number of data in Power Ramp for the
query command.
The maximum number is 1325.
The range of a value is from
-9.250
to
156.250
Bits in GMSK.
The step is 0.125 Bits step and the number of data is 1325 in
GMSK.
The range of a value is from
-27.750
to
468.750
Bits in 8PSK.
The step is 0.375 Bits step and the number of data is 1325 in
8PSK.
The set value by this command is a common value to TCH1 to
TCH6 Manual Test and TX Analyzer.
Power Ramp Trace
TXANalyzer:RESults:RAMP:TRACe?
This query command returns a value of the trace data in Power
Ramp.
The range of a value is from
-9.250
to
156.250
Bits in GMSK.
The step is 0.125 Bits step and the number of data is 1325 in
GMSK.
The range of a value is from
-27.750
to
468.750
Bits in 8PSK.
The step is 0.375 Bits step and the number of data is 1325 in
8PSK.
The start point is set by the command
TXANalyzer:RESults:RAMP:TRACe:STARt.
The number of data is set by the command
TXANalyzer:RESults:RAMP:TRACe:NUMber.
•
<int1>
is
0
if measurement result is within the
measurement range, or
1
if measurement result is out of
range or the Access Probe Power is not measured.
•
<real2>
is the start point of data.
•
<real3>
is the number of step.