Programming Command Reference
5
N9360A Multi UE Tester GSM Programming Manual
5-155
•
<int 3>
(2 digits) is the measured RX Level factor ranging
from
0
(<-110 dBm) to
63
(>–48 dBm). If measurement result
is out of range or the RX Level is not measured, the value
63
is returned.
BER BS Level
TESTs:MANual:TXAMplitude:RESults:MARKer:POINt
TESTs:MANual:SENSitivity:LEVel <real>
TESTs:MANual:SENSitivity:LEVel?
This command sets the amplitude (BS Level) for Sensitivity BER
tests for each radio system set by the
command
. The allowable range is from
-110.0
to
-50.0
(dBm) in
0.1
steps. At preset, this is set to
-102.0
(dBm) for GSM850, GSM900 and PCS1900, or
-100.0
(dBm) for
DCS1800.
This query command returns the set amplitude (BS Level) for
Sensitivity BER tests with real ranging from
-110.0
to
-50.0
in dBm.
Sensitivity BER and FER Measurement (GSM mode, Testloop Mode
A)
TESTs:MANual:RESults:SENSitivity?
This query command returns the Sensitivity measurement
result values
<int 1>
,
<int 2>
,
<real 3>
,
<real 4>
and
<real 5>
, when GSM mode is selected and the Test Loop Mode
is set to
A
. Each value is separated by a comma.
•
<int 1>
(1 digit) is
0
if measurement results are within the
measurement range, or
1
if any measurement result is out of
range or the Sensitivity is not measured.
N O T E
If the mode is set to GPRS or EGPRS, this query command is invalid.