EGPRS (MCS5-9): <real 3> is the marker position in bits ranging
from
27.75
to
468.75
when the zoom
function is set to
Off
, or from
-24.00
to
6.00
and
435.00
to
465.00
when the zoom
function is set to
On
.
•
<real 4>
(up to 5 letters) is the measured power level at the
marker position. If measurement result is out of range or the
Power Ramp is not measured, the value
-99.9
is returned.
Power Ramp Marker Position Control
TESTs:MANual:TXAMplitude:RESults:MARKer:POINt
UP|DOWN
TESTs:MANual:TXAMplitude:RESults:MARKer:POINt
<real>
TESTs:MANual:TXAMplitude:RESults:MARKer:POINt?
GSM/GPRS/EGPRS (MCS1?4): This command moves the marker
position UP or DOWN in steps, or
by <real> in bits ranging from
-9.25
to
156.25
when the zoom
function is set to
Off
, or from
-8.00
to
2.00
and
145.00
to
155.00
when the zoom function is
set to
On
.
EGPRS (MCS5?9): This command moves the marker position by
<real> in bits ranging from
-27.75
to
468.75
when the zoom function is set to
Off
, or from
-24.00
to
6.00
and
435.00
to
465.00
when
the zoom function is set to
On
.