Programming Command Reference
5
N9360A Multi UE Tester GSM Programming Manual
5-173
Power Ramp Measurement
TXANalyzer:RESults:RAMP?
This query command returns the Power Ramp measurement
result values
<int 1>
and
<int 2>
. Each value is separated by
a comma.
•
<int 1>
(1 digit) is
0
if measurement result is within the
measurement range, or
1
if measurement result is out of
range or the Power Ramp is not measured.
•
<int 2>
(1 digit) is the test result for the
0
(pass) or
1
(fail).
Power Ramp Graph Zoom
TXANalyzer:RESults:RAMP:ZOOM ON|OFF|1|0
TXANalyzer:RESults:RAMP:ZOOM?
This command sets whether to activate the zooming function on
the Power Ramp graph screen. The parameter is
ON
(1) or
OFF
(0). If the parameter is set to
ON
, the rising and trailing edges
are expanded. At preset, this is set to
0
.
This query command returns the selected zooming function on
the Power Ramp graph screen with integer value of
1
or
0
. The
meaning of the return value is same as that of the command
parameter.
Power Ramp Marker Measurement
TXANalyzer:RESults:RAMP:MARKer?
This query command returns values at marker position on the
Power Ramp graph screen by values
<int 1>
,
<int 2>
,
<real
3>
and
<real 4>
. Each value is separated by a comma.
•
<int 1>
(1 digit) is
0
if measurement result is within the
measurement range, or
1
if measurement result is out of
range or the Power Ramp is not measured.
•
<int 2>
(1 digit) is the test result for the
0
(pass) or
1
(fail).