5-154
N9360A Multi UE Tester GSM Programming Manual
5
Programming Command Reference
Command Reference for Sensitivity/RX Quality/RX Level Measurement
Screen
The SENSitivity commands are used to set or get the controls
and parameters associated with the Sensitivity/RX Quality/RX
Level measurements and the pass or fail tests.
Power Control Level
TESTs:MANual:SENSitivity:PLEVel <int>
TESTs:MANual:SENSitivity:PLEVel?
This command sets a value of Power Control Level for each
radio system set by the
command described on
. The allowable range and preset values are as
follows:
RX Level Measurement
TESTs:MANual:RESults:LEVel?
This query command returns the RX Level measurement result
values
<int 1>
,
<int 2>
and
<int 3>
. Each value is
separated by a comma.
•
<int 1>
(1 digit) is
0
if measurement result is within the
measurement range, or
1
if measurement result is out of
range or the RX Level is not measured.
•
<int 2>
(1 digit) is the test result for the
0
(pass) or
1
(fail).
N O T E
If the mode is set to GPRS or EGPRS, the commands related to RX Quality
and RX Level measurements and their pass/fail tests results are invalid.
Table 5-39 Power Control Level Allowable Range and Preset Value
RFORmat
Allowable Range
Preset Value
GSM850
2 (+39 dBm) to 19 (+5 dBm)
5 (+33 dBm)
GSM900
2 (+39 dBm) to 19 (+5 dBm)
5 (+33 dBm)
DCS1800
0 (+30 dBm) to 15 (+0 dBm), or 29 (+36 dBm) to 31 (+32 dBm)
0 (+30 dBm)
PCS1900
0 (+30 dBm) to 15 (+0 dBm), or 30 (+33 dBm) to 31 (+32 dBm)
0 (+30 dBm)