Maintaining the Purged Packed Inlet
To Install a Packed Metal Column
Maintaining Your GC
93
To Install a Packed Metal Column
1
Gather the following:
•
7/16-inch, 9/16-inch, and 1/2-inch wrenches
•
Lint-free gloves
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
3
Prepare the packed metal column. See
“To Install Ferrules on a Packed Metal Column”
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
4
Install the 1/8-inch or 1/4-inch packed column inlet adapter, if necessary. See
an Adapter on the Purged Packed Inlet”
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
5
Attach the column to the inlet adapter. Finger-tighten the nut.
6
Tighten the nut an additional 1/4 turn with a wrench (for a 1/8-inch column) or an
additional 3/4 turn (for a 1/4-inch column).
Use two wrenches, one on the column nut and the other on the adapter, to prevent the
adapter from rotating.
7
Press
Settings > Configuration > Columns
, then select
Columns 1
through
6.
Select
Column Type > Packed
, then choose your
Inlet
and
Outlet Connection
to identify the inlet
and detector to which the column is attached.
8 Scroll
down to manually set your column temperatures.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an
explosion hazard.
9
Condition the column, if necessary. See
“To Condition a Packed Column”
10
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
11
If required, install a packed column adapter onto the detector column fitting. See
Install a Packed Column Adapter onto a Detector”
12
Place a nut and ferrule onto the packed column.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...