![Agilent Technologies 8890 Maintaining Manual Download Page 62](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678062.webp)
5
Maintaining the Split/Splitless Inlet
To Change the Liner and O-Ring on the Split/Splitless Inlet
62
Maintaining Your GC
To Change the Liner and O-Ring on the
Split/Splitless Inlet
1
Gather the following:
•
Replacement O-ring, see
“Consumables and Parts for the Split/Splitless Inlet”
•
Replacement liner
•
Tweezers
•
Wrench, hex for changing septum (optional)
•
Wrench, capillary inlet (optional)
•
Lint-free gloves
2
Launch the GC maintenance wizard:
Maintenance > Inlets,
select the inlet, then select
Perform Maintenance > Replace Liner > Start Maintenance.
The wizard will walk through
the steps needed to replace the liner and O-ring (those same steps are repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
3
Slide the locking tab forward (counterclockwise). Lift the insert assembly straight up and
away from the inlet to avoid chipping or breaking the liner.
4
Loosen the O-ring from the sealing surface with tweezers.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...