5
Maintaining the Split/Splitless Inlet
To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet
68
Maintaining Your GC
Figure 23.
Slide trap assembly to expose the filter
6
Remove the old filter cartridge and two O-rings.
7
Verify the new O-rings are seated properly on the new filter cartridge.
8
Install the new filter cartridge then reassemble the trap using your fingers to fully tighten
the knurled nut. The nut should thread easily.
If the knurled nut does not thread easily, loosen, reassemble the 2 parts of the trap, and
thread again. Do not force it.
9
The GC maintenance wizard will perform checks at the appropriate times, including
Leak &
Restriction
tests, and will automatically reset the maintenance counters.
10
Install the pneumatics cover.
11
Select
Finished
,
then select
OK
to exit the GC maintenance wizard
.
Knurled nut
Filter and O-ring
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...