13 Maintaining the NPD
To Maintain the NPD Collector, Ceramic Insulators, and Jet
222
Maintaining Your GC
8
Remove the screws that secure the interconnect cover, then lift up the cover and set aside.
9
Remove the screws that secure the electrometer, then lift up and set aside.
10
Pull the electrometer away from the detector to free the interconnect. Turn the
electrometer to the right to obtain working space. Be careful not to touch or bend the
spring. Be careful not to lose the EMI suppressor.
11
Remove the large metal C-ring and the upper ceramic insulator if they were not attached to
the lid.
12
Remove the collector. If the detector is operated at high temperatures, the collector parts
may stick inside the detector. Gently push and wiggle them to break the seal.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...