5
Maintaining the Split/Splitless Inlet
To Install a Capillary Column with the Split/Splitless Inlet
56
Maintaining Your GC
10
Thread the column nut into the inlet but do not tighten.
11
Adjust the column position so that the septum contacts the bottom of the column nut.
Finger-tighten the column nut until it begins to grip the column.
12
Tighten the column nut an additional 1/4 to 1/2 turn with a wrench so that the column
cannot be pulled from the fitting with gentle pressure.
13
Plug in the new column’s Smart ID key. See
.
Figure 21.
Insert Smart ID key
14
Configure the new column.
15
Condition the column per the manufacturer’s recommendation. See
4
%
6 mm
Smart ID key
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...