11 Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
192
Maintaining Your GC
7
Remove the gasket from the bottom of the assembly, if necessary.
8
Remove the FID ignitor castle.
a
Loosen the collector nut.
b
Remove the collector nut and the spring washer.
c
Lift the castle out of the collector housing. When removing the castle, some of the
collector parts may be attached. Set these on a clean cloth to protect from scratches
or dirt.
9
If only replacing the FID castle, skip to
for reassembly.
10
Remove the collector and insulators.
a
If needed, remove the collector and upper insulator from the FID housing. The lower
insulator may come out with the collector, but often remains in the FID housing. Place
the parts on a clean cloth.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...