11 Maintaining the FID
To Replace an FID Jet
Maintaining Your GC
189
10
Install the castle assembly.
11
Select
Finished
,
then select
OK
to exit the GC maintenance wizard
.
12
Reset the EMF counter.
13
Attach the capillary column to the detector.
a
Install the column in the detector. See
“To Install a Capillary Column in the FID”
b
After the column is installed at both inlet and detector, establish a flow of carrier gas
and purge as recommended by the column manufacturer.
c
Check the FID leakage current. See
“To Check the FID Leakage Current”
d
Bakeout the detector. See
e
Restore the analytical method.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
f
Allow the oven, inlet, and detector to equilibrate at operating temperature, then
retighten the fittings.
14
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...