11 Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
196
Maintaining Your GC
17
Lower the collector assembly into the housing.
18
Insert the three screws and tighten (to 18-inch-pounds).
19
Connect the ignitor extension cable.
20
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
21
Verify assembly:
a
Check the FID leakage current. See
“To Check the FID Leakage Current”
b
Bakeout the detector. See
c
Check the FID baseline. See
22
Reset the EMF counters.