![Agilent Technologies 8890 Maintaining Manual Download Page 160](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678160.webp)
9
Maintaining the PTV Inlet
To Replace the Filter in the Split Vent Trap for the PTV Inlet
160
Maintaining Your GC
To Replace the Filter in the Split Vent Trap for the
PTV Inlet
1
Gather the following:
•
New split vent filter cartridge, see
“Consumables and Parts for the Programmable
Temperature Vaporization Inlet”
2
Launch the GC maintenance wizard:
Maintenance > Inlets,
select the PTV inlet, then select
Perform Maintenance > Replace Split Vent Trap > Start Maintenance.
The wizard will walk
through the steps needed to replace the filter (those same steps are repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
WARNING
The split vent trap may contain residual amounts of any samples or other chemicals you
have injected into the GC. Follow your company’s safety procedures for handling these types
of substances while replacing the trap filter cartridge.
3
Remove the
“To Remove the Pneumatics Cover”
on page 19 (top, back of GC).
4
Completely loosen the knurled nut that secures the split vent trap in place as shown in
.
Figure 34.
Loosen knurled nut
5
Slide the trap assembly back from the guided mounting bracket and tilt up to expose the
filter as shown in
.
Knurled nut secured
Knurled nut loosened
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...