9
Maintaining the PTV Inlet
To Replace the Septumless Head PTFE Ferrule on the PTV Inlet
Maintaining Your GC
153
12
Observe the inlet
Total Flow
. Tighten the guide cap until the
Total Flow
stops decreasing
(typically near 8 mL/min).
13
Remove the syringe from the inlet and press
.
14
Restore the analytical method.
15
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.