Maintaining Capillary Columns
To Remove the Ferrule From a CFT Fitting
Maintaining Your GC
47
To Remove the Ferrule From a CFT Fitting
Loosen and remove the internal nut. If the ferrule does not release from the fitting, do not try to
pull the column free. Insert a pointed object (such as a thumb pin or a paper clip) into the
ferrule release hole, as
shown
, and press firmly. You will hear a click as the ferrule
releases.
Figure 19.
Release the ferrule (Ultimate Union fitting shown)
Ferrule release holes