![Agilent Technologies 8890 Maintaining Manual Download Page 154](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678154.webp)
9
Maintaining the PTV Inlet
To Change the Septum on the PTV Inlet
154
Maintaining Your GC
To Change the Septum on the PTV Inlet
1
Gather the following:
•
Replacement septum, see
“Consumables and Parts for the Programmable
Temperature Vaporization Inlet”
•
5/8-inch wrench
2
Launch the GC maintenance wizard:
Maintenance > Inlets,
select the inlet, then select
Perform Maintenance > Replace Septum > Start Maintenance.
The wizard will walk
through the steps needed to replace the septum (those same steps are repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
3
Remove the septum retainer nut or Merlin cap. If the septum head begins to turn during
removal, support it manually while removing the cap.
4
Use tweezers to remove the septum or Merlin Microseal from the retainer nut. Do not
gouge or scratch the interior of the septum head.
5
Firmly press the new septum or Merlin Microseal into the fitting. The metal parts side of
the Merlin Microseal should face down (toward the oven).
6
Replace the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum
retainer nut until the C-ring is about 1 mm above the nut.
CAUTION
Overtightening the septum nut can cause contamination.
7
The GC maintenance wizard will perform checks at the appropriate times, including
Leak &
Restriction
tests, and will automatically reset the maintenance counters.
8
Select
Finished
,
then select
OK
to exit the GC maintenance wizard
.
9
Restore the analytical method.
Merlin Microseal and cap
Standard septum
1 mm
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...