9
Maintaining the PTV Inlet
To Replace the Inlet Adapter for the PTV Inlet
Maintaining Your GC
159
5
Remove the inlet adapter and discard the old silver seal.
6
Insert a new silver seal into the adapter and finger-tighten the adapter onto the inlet.
Tighten an additional 1/16 to 1/8 turn with a wrench; overtightening will damage the inlet.
7
Install the column. See
“To Install a Capillary Column with the Programmable
Temperature Vaporization Inlet”
8
Check the adapter for leaks.
9
Reset the silver seal EMF counter.
10
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.
11
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
12
Restore the analytical method.
Silver seal
Graphpak inlet adapter
Column nut
Ferrule for Graphpak adapter
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...