![Agilent Technologies 8890 Maintaining Manual Download Page 233](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678233.webp)
14 Maintaining the ECD
To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter
Maintaining Your GC
233
To Replace the ECD Fused Silica Indented Mixing
Liner and Install the Makeup Gas Adapter
1
Gather the following:
•
Fused silica indented mixing liner, see
“Consumables and Parts for the ECD”
•
1/4-inch Swagelok nut
•
1/4-inch Vespel/graphite ferrule
•
9/16-inch wrench
•
Methanol
•
Lint-free gloves
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready to continue (all
zones cooled to safe handling temperatures).
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3
Remove the column from the makeup gas adapter.
CAUTION
To prevent damage avoid flexing/bending the tubing on the makeup gas adapter.
4
Remove the makeup gas adapter.
a
Loosen the adapter nut with a wrench and slide out the makeup gas adapter from the
ECD. Remove the ferrule.
The makeup gas adapter will remain attached to the supply tubing and hang
suspended in the oven.
b
Adjust the adapter’s position so that maintenance can be performed on the adapter
easily and without obstruction.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...