4
Maintaining Capillary Columns
Preparing the column installation to the inlet
42
Maintaining Your GC
8
Using the 1/4-inch open-ended wrench, remove the internal nut from the swaging tool and
column.
Figure 11.
Properly-trimmed column and ferrule
Preparing the column installation to the inlet
This procedure is used to attach a capillary column to a splitter or switch or an Ultimate Union.
1
Gather the following:
•
UltiMetal Plus Flexible Metal ferrules. See
.
•
Swaging tool (G3440-80218)
•
Two 1/4-inch open-end wrenches
•
Column cutting tool (5181-8836)
•
Column nut
•
Lint free gloves
2
Check the column end. It should be square and free of cracks. If needed, trim it. See
.
Figure 12.
3
Pass the column end through the column nut, the ferrule, and the swaging tool, as shown
in
.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...