Maintaining the Purged Packed Inlet
To Clean the Septum Seat in the Purged Packed Inlet
Maintaining Your GC
85
12
On the GC touchscreen select
Maintenance
>
Inlets
>
Septum injections
,
then select
Reset
Counter
.
13
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.
14
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
15
Restore the analytical method.
1 mm